Make sure that the arrow points towards the pump.

2.

Switch the pump on and adjust the Low Flow adapter to the required

flow rate using the screw mounted in the side.

3.

Connect the pump to the flow meter and set the flow rate to

1.2 - 1.5 l/min.

4.

Switch the pump on and adjust the Low Flow adapter to the required

flow rate using the screw mounted in the side. The flow may be

calibrated using a low flow calibrator or a Bios Dry Cal Calibrator.

5.

Remove and replace the sorbent tube.

6.2.2 Inserting Tubes into Their Holders

1.

Check that the flowmeter used is operating at the correct rate by

using a spare sampling tube

2.

Break off both ends.
A tube cutter is available from Casella CEL.

3.

Place one end in the tube connected to the pump, the other end

connects to the flowmeter.
To determine which way round the tube should go, look for the arrow

painted on the side. The arrow indicates the direction of flow and

must point towards the pump.

4.

Using a Dry Flow flowmeter or Bubble flowmeter, adjust the flow

rate according to the instructions, to suit the specific application.

With the flow rate set, remove the tube from the tube holder and replace

with a clean one, making sure that the arrow points towards the pump.
When sampling is complete, the sealing caps must be replaced before the

tube is sent for analysis.

02058

Flow

Apex Pump

Constant Pressure Controller

Low Flow Adaptor

Sorbent Tube

in Tube Holder

PROG.

SET FLO

oC

L/min

RESET

CAL

I.R.

TEMP 19.48

2.5

Flow

Figure 16: Low flow measurement arrangement using a CPC and Low Flow

Adaptor