Prog. set flo oc l/min reset cal i.r – Casella CEL Sampling heads for personal dust sampling User Manual

Page 17

Advertising
background image

6.1.3 Calibration at High Flow For Impingers and Bubblers

Figure 14 shows the arrangement for calibration of impingers and bubblers.
Please refer to the correct flow rate for the impinger / bubbler before

calibrating the flow.

6.2

Calibration at Low Flow

(Gas and Vapour Sampling)

6.2.1 Calibration at Low Flow Using

CPC / Low Flow Adapter

Sampling can be carried out using either an

Apex pump (0.5 - 5.0 l/min) or a Vortex Low

Flow pump (5 - 200 ml/min).

When using an Apex pump for low

flow applications, a Low Flow kit is

necessary. Two kits are available, both with

Low Flow Adapter (P109026) and Constant

Pressure Controller (P109027) as shown in

Figures 15 and 16. One kit includes the

small Tube Holder P109023 while the other

has the large Tube Holder P109024.
1.

Connect the pump to the CPC and

low flow adapter as shown in the

figure. Break both ends off of a

sorbent tube, and place between

the adapter and bubble flowmeter.

02059

Flow

Flow

Apex Pump

Impinger / Bubbler

Optional items for use
ONLY in low flow
applications

Carbon Trap

P109018

PROG.

SET FLO

oC

L/min

RESET

CAL

I.R.

TEMP 19.48

2.5

Constant Pressure

Controller P109027

Low Flow Adaptor

P109026

Figure 14: Calibration of impingers and bubblers

Calibration

Figure 15: Low flow kit

P109032 with large tube holder

Page 17 of 24

SAMPLING HEADS AND ACCESSORIES
Users Guide