Casella CEL DryCal DC-Lite flow calibrator User Manual

Page 2

Advertising
background image

Table of Contents

1.0

DC-Lite Features

2

2.0 Unpacking

Checklist

3

3.0

General Description

3

4.0

Theory of Operation

3–4

5.0

Operating Instructions

4

5.1

DC-Lite Button Panel

4

5.2

Turning the Power On

5

5.3

Disabling & Re-Enabling the 5-Minute Auto-Shutoff Feature

5

5.4 Taking

Readings

5–6

5.5

Resetting the Averaging Sequence

6

5.6

Printing

6–7

5.7

Stop & Reset

7

5.8

Resetting a Printed Sequence

7

5.9

Printing to a PC

8

5.10

Hard Reset Button

8

6.0

Battery System

8

6.1

Charging the Battery

9

6.2

Battery Maintenance & Storage

9

7.0

Isolating the DryCal from Other Instruments

9–10

7.1

Use with Instruments that Contain Internal

Mass Flow Controllers (MFCs)

10

7.2

Use with Personal Air Samplers

10

7.3

Calibrating Rotameters

10

7.4

Use with Magnehelic Manometers

10–11

8.0

Maintenance, Quality Assurance

11

8.1

Maintenance 11

8.2

Leak-Test Procedure

11–12

8.3

Air Containing Particulates

12

8.4

Return Authorization

12

8.5

Shipment

12

8.6

Long-Term Storage

12–13

8.7

Calibration 13

9.0

Calibration Statement

13–14

10.0

Limited Warranty

14

DryCal

®

DC-Lite Specifications

Size 5" x 5" x 2.75" 127 mm x 127 mm x 70 mm

Weight 42 oz 1200 g

Flow Ranges | Air Flow Accuracy Specifications based on averaged readings.

Lower limit is based on self-tested maximum leakage.

Model

Optimum Flow Range (±1%)

Extended Flow Range

L

10–500 ml/min

1 ml/min–500 ml/min

ML

50 ml/min–2 L/min

5 ml/min–5 L/min

M

100 ml/min–7 L/min

10 ml/min–12 L/min

MH

200 ml/min–20 L/min

20 ml/min –20 L/min

H

500 ml/min–30 L/min

50 ml/min–30 L/min

Contact Bios for extended flow range specifications, or visit our website

at www.biosint.com/products/dclite_models.htm

Battery System 6V rechargeable, sealed lead-acid, 6-8 hours typical operation

AC Battery Charger | Power Adapter Wall-mounted, single-station charge,

input: 100 to 120 VAC, 60 Hz., output: 12 VDC. Optional input: 200 to 240 VAC,

50 Hz., output 12 VDC.

Operating Modes Single reading, 10 readings, or auto-mode.

Temperature Range 0–55 °C

Humidity Range 0–70% non-condensing

Printer Port Standard parallel (Not compatible with printers that require

Microsoft

®

Windows

)

Warranty Product, 1 year; battery, 6 months

The annual recalibration program offered by Bios is elective and is not included

as a warranty item. All specifications are subject to change.

Please contact Bios or visit our web site at www.biosint.com for the most

current information.

Advertising