2.0 Unpacking Checklist

Your DryCal DC-Lite has been packaged with care and includes all components

necessary for operation. Please take a moment to check that you have received

the following items. If you believe you have not received a full shipment or have

any other questions, please contact Bios immediately.

Your DryCal DC-Lite Includes

Single-Station Battery Charger

Tubing Kit

Additional High Flow Tubing (with DCL-MH and DCL-H only)

Certificate of Calibration

Instruction Manual

Registration Card

3.0 General Description

The DryCal DC-Lite is a field-portable primary flow calibrator used for industrial

hygiene, environmental and laboratory flow measurement applications.

The DC-Lite uses patented dry piston technology and infrared sensors to obtain

volumetric flow rates quickly and accurately.

Housed in a small, sturdy case, each unit employs a variety of popular user con-

veniences such as push-button read and auto-read functions, a large alphanumeric

display, battery level indicator, 5-minute automatic shut-off and a parallel printer

port for data-logging.

4.0 Theory of Operation

The DryCal DC-Lite can be used to measure air flow rates for either a vacuum flow

source (connected to the outlet port) or a pressure flow source (connected to the

inlet port). Before a reading is initiated, or between readings, a computer-directed

valve performs a bypass function. This allows the air to pass through the DryCal

valve, bypassing the flow-measuring cell which is then able to reset.

As a reading is initiated (by pressing the Read button) the internal valve closes

and the flow source evacuates or pressurizes the air in the flow-measuring cell. The

piston rises at the rate of evacuation or pressurization. A precision encoder system

provides two finely collimated light beams with a known distance between the

beams.

After a suitable acceleration interval the piston breaks the first infrared light beams

as it passes. The flow reading is completed when the second infrared beam

3

2

1.0 DC-Lite Features

Front and Side

Back and Side

Cell Viewing Window

Charging Jack

Inlet Port
Connect tubing here for
pressure applications

Outlet Port
Connect tubing here for
suction applications

Parallel Printer Port

Reset Button