6.1 Charging the Battery

Before using your DryCal DC-Lite, be sure that the battery system has been fully

charged to ensure that unit will perform without interruption. Using the DC-Lite

with a low battery will not affect the product’s accuracy.

The DC-Lite is equipped with a battery indicator that provides battery charge

indication at three levels. When the battery indicator on the display is empty the

unit will continue to operate for a limited period of time before shutting itself off.

To Charge the DC-Lite

To view the actual charging status during the charging period, disconnect the

battery charger and wait 3–5 minutes. When the indicator is solid black the battery

is fully charged. Bios recommends leaving the DC-Lite on charge when not in use

to prevent battery degradation.

1

Connect only the appropriate Bios 12VDC charger, provided with the DC-Lite

calibrator, into a standard wall outlet.

2

Insert the charger barrel plug into the charging jack located on the right side

of the DC-Lite housing above the inlet and outlet air bosses. A green Charge

LED will illuminate while the unit is charging. Full charge takes 8 to 12 hours,

and the DryCal can charge while being used.

6.2 Battery Maintenance & Storage

The DC-Lite’s lead-acid battery will not exhibit the memory effect common to

nickel-cadmium batteries. It may be left on charge for an indefinite time period

without damage.

Long-term storage without charging can damage the battery pack,

therefore if the DC-Lite cannot be left charging continuously, it should be

fully charged at least once every three months and should be placed in

storage only after achieving a full charge.

7.0 Isolating the DryCal from Other Instruments

The DryCal DC-Lite will mimic the flow source being used. Therefore, if the flow

source exhibits air flow pulsation, Bios recommends the use of an isolation device.

Use of a 25mm, 0.8m filter cassette makes a suitable load for most flow rates used

in industrial hygiene applications. This method stabilizes variations in flow due

to the slight pulsation caused by the stroke of the pump’s piston.

In addition, when taking flow readings with the DryCal DC-Lite, an internal valve

closes, placing an insertion pressure spike into the flow stream. Generally, the

9

5.9 Printing to a PC

Bios International offers a parallel-to-serial converter kit, part PSC-1, that allows

the information from a DC-Lite to be printed to a computer via the HyperTerminal

utility included with Microsoft

®

Windows

TM

.

This information can be imported into many commonly used spreadsheet

programs, such as Microsoft Excel or Quattro Pro. The DryCal parallel-to-serial

converter kit includes everything you will need to print flow readings from your

DryCal to a Windows-based PC.

Bios International only guarantees compatibility with parallel-to-serial converters

purchased through Bios International. Bios International does not offer technical

support on serial port configuration. For assistance with determing the correct

Com Port number or port configuration, please contact your IT professional.

5.10 Hard Reset Button

If for any reason the DC-Lite does not respond to push-button commands, it may

be necessary to reset the instrument. For this purpose there is a white recessed

button on lower right side of the back panel near the parallel printer port. The but-

ton resets the unit back to the initializing screen and the printer setting will revert

to the “Off” position. Before resetting, be sure to remove the printer cable from

the back of the DC-Lite. Failure to do so will result in the display reading “Nexus

Control.” If this occurs, remove the printer cable and reset the unit again.

6.0 Battery System

The DryCal DC-Lite is powered by an internal lead-acid battery. The battery will

power the instrument for 6–8 hours of continuous use and has a typical service life

of approximately 2–5 years, depending on use. The DC-Lite provides a convenient

5-minute automatic shut-off feature to extend battery life. Use of a printer does not

affect the battery life.

The DC-Lite can be charged by the Bios single-station charger when plugged into

a standard 115V AC power source outlet (220V AC optional). Provided that the

battery has sufficient charge to operate the DC-Lite, the DC-Lite can be charged

indefinitely using the AC wall adapter provided.

Although the DC-Lite may be plugged into AC power, if the battery is exceptionally

weak the DC-Lite may not function. Please read all setup and charging instructions

indicated in this manual before using equipment.

8