7 calibration, 0 calibration statement, 3 air containing particulates – Casella CEL DryCal DC-Lite flow calibrator User Manual

Page 8: 4 return authorization, 5 shipment, 6 long-term storage

Advertising
background image

charged at least once every three months.

Always store DryCal calibrators in a clean, dry environment and recharge

the unit prior to use after long-term storage.

8.7 Calibration

As a quality assurance measure, Bios recommends annual calibration of all meas-

urement instruments, although how often you have your DryCal calibrated is

an internal quality control decision. The determining factors are whether the unit

passes the internal leak-test, quality system requirements if applicable, and

the conditions in which the unit is used. Units used in a laboratory setting may

require calibration less frequently than a unit that is used in a dusty environment.

The annual calibration program is an elective and is therefore not included as

a warranty item. “As received” flow test data and expedited “48 hour” turnaround

service are also available at an additional cost. Please contact the factory for more

information on available calibration services and pricing.

Calibration Includes

Cleaning (if required)

Valve adjustment (if required)

Battery capacity test

Internal computer program upgrade as necessary

Mechanical upgrades as necessary

Dynamic Performance Test

NIST-Traceable Calibration Certificate

9.0 Calibration Statement

The DC-Lite is dynamically tested by comparing it to a Laboratory Standard pri-

mary piston prover of much higher accuracy (±0.25%), but of similar operating

principles. Flow generators of ±0.01% stability (included in prover accuracy) are

used for the comparison. Use of provers of similar construction to the devices

under test assures the validity of the flow generator as a transfer standard.

The primary Laboratory Standards are qualified by direct measurement of their

dimensions (diameter, length of measured path, time base) against NIST-traceable

gauges and instruments. A rigorous analysis of their accuracy in accordance with

the International Guide to Uncertainty in Measurements has been performed,

assuring their traceable accuracy. Test procedures assure temperature matching of

the Laboratory Standards to the devices under test.

The calibration dates of the laboratory standards for each parameter (diameter,

13

valve will close. Return the unit to an upright position and it will time the

descent of the piston.

3

Place the DC-Lite on a flat, vibration-free surface.

4

Observe the location of the piston to ensure that it is at the top of the cell

when the test begins (the test may take as long as 15–20 minutes). If the

test is completed successfully, the display will read: “Test OK Push Read.”

5

Push the Read button as directed and the internal valve will open

and the piston will fall.

6

Repeat the test with the leak-test tubing accessory connected to the

lower (outlet) air boss. If the unit fails the Leak-test, the display will

read: “Maintenance Reqd Push Read.”

8.3 Air Containing Particulates

As of January 1, 2001, the DryCal DC-Lite comes standard with either a 5-micron or

30-micron inlet filter inside the inlet fitting (depending on model ordered).

Additionally, all older DC-Lites sent in for calibration will be retrofitted with new

style inlet filters, free of charge. However, air containing cigarette smoke or other

excessive dust and particulates should be additionally pre-filtered. An additional

particulate filter, part AF-516, is available for this purpose. The filter should be

placed ahead of the DryCal in the flow stream, on the inlet side.

8.4 Return Authorization

Prior to returning your DryCal for repair or recalibration, please contact Bios

International for technical support, troubleshooting assistance and an RMA

number if necessary.

You can telephone Bios at (800) 663 4977 or (973) 492 8400, or send an email to

service

@

biosint.com.

8.5 Shipment

When shipping the DryCal DC-Lite please ensure that the packaging is adequate

to protect the instrument. When possible the DC-Lite should be shipped in

the original packaging. Bios International Corp. is not responsible for damage that

occurs during shipment.

8.6 Long-Term Storage

DryCal calibrators can remain on charge until needed without causing damage

to the battery. If the DryCal is stored for long periods of time the battery should be

12

Advertising