background image

6 | SETTINGS MENU

The following controller modes can be selected:

selection

description

analog input

Setpoint via analog input

bus/rs232

Setpoint via bus or RS232 (BRIGHT)

rs232

Setpoint via RS232 only (BRIGHT)

fb ana slave

Setpoint is factor of measure of

master on the bus, factor is set by

analog input.

analog slave

Setpoint is factor of analog input

fb slave

Setpoint is factor of measure of

master on the bus

valve open

Setpoint neglected, valve fully open

valve close

Setpoint neglected, valve closed

controller idle

Setpoint neglected, valve stays in

actual position

setpoint 100%

Setpoint is 100 perc.

setpoint 0%

Setpoint is 0 perc.


Use or to select one of these control modes. Press

to cancel selection and exit the edit mode

without changes
When the required control mode is selected, press

to

confirm.

6.8 Master slave settings

When you select one of the three slave modes the ‘master

slave’ option will be added to the controller settings

screen.

CANCEL

SET

9.17.048

Page 21

Advertising