p20

Parameter
controlled

Limits

Function

Key sequence to
enter “Set” mode

Temperature
Variation Limit

3 – 9 °C

This setting allows you to
determine the amount of
variation allowed in the water
bath temperature before
an alarm is triggered. For
example when set to 3 °C, the
temperature alarm will trigger
if the water bath temperature
becomes 3 °C above or
below the desired water bath
temperature.

SET + ▲
(simultaneously)

Upper Temperature
Limit

50 – 110 °C

Triggers an alarm when the
water bath temperature reaches
the setting.

SET + ▼
(simultaneously)

Set Water Bath
Temperature

-10 to
+110 °C

Sets the desired internal
or external water bath
temperature.

SET

Change Water Bath
Temperature

Change the water bath
temperature while in use.

ADJ

Setting Temperature Parameters

The following temperature parameters can be controlled with the RCB20-PLUS:
• Temperature Variation Limit
• Upper Temperature Limit
• Water Bath Temperature

Table 1. Parameters and Controls for RCB20-PLUS