p21

Water bath temperature setting

Refrigeration system function

≤ 25 °C

Refrigeration system will cool.

25 – 30 °C

The refrigeration system will remain powered but
will not cool.

≥ 30 °C

The refrigeration system powers will automatically
shut off. Once this happens, the user will not be able
to turn the refrigeration unit back on for 3 minutes.

Table 2. Refrigeration System Temperature Control Limitations

Note: This table assumes
the unit is being used at
room temperature, 20 °C.
If using the water bath in a
cold room, 16 °C should be
subtracted from the water
bath temperature settings.