Getting started with cart – Metro Starsys Mobile Power User Manual

Page 4

Advertising
background image

4

GETTING STARTED WITH CART

Unpacking

The Starsys Mobile Power cart will arrive fully assembled and fully functional at the customer’s site (most accessories do
not ship assembled to the cart. After cutting the strapping bands, lift the top of the box over the cart. In order to avoid
any injury, two (2) people should lift the cart from the padding blocks.

Power Supply Preparation

CAUTION

: Before placing the Li-Nano Power Supply into service on a Starsys cart, the Li-Nano batteries must be

connected. These batteries have been disconnected prior to shipment due to USDOT and IATA regulations.

Connecting the Power Supply Batteries

CAUTION

: Only qualifi ed service personnel should perform the following procedure for connecting the Power Supply

Batteries.

1. The AC cable must be disconnected from the wall outlet as shown in Figure 2.

Figure 2: Disconnect AC Cable

2. The Normal Operation/OFF switch of the Power Supply must be in the Standby (down) position as shown in Figure 3.

Figure 3: Power Supply Switch Down to Standby

3. Remove screw from each side of power supply that hold the battery connector access panel as shown in Figure 4.

Figure 4: Power Supply Rear View

4. Rotate the battery connector access panel down by gripping the tab on either side as shown in Figure 5.

Figure 5: Battery Connector Access Panel

REMOVE SCREW

REMOVE SCREW

Advertising