WARNING

SYSTEM MAY BE UNDER HIGH PRESSURE. When removing the flow meter, be sure to hold

the electronics enclosure firmly by hand before slowly loosening the positioning clamping

nut. Failure to do this will allow the pressure in the pipe to suddenly and rapidly force the

meter from the pipe causing serious injury. The meter could also be damaged or break apart

causing a break in the water seal with the resultant loss of large amounts of water. The hand

effort required to hold the meter will be 0.11 times the pipe pressure.

!

Step 1:
Slowly loosen the position
clamping nut while

maintaining a firm grip on the
enclosure to counteract the
effect of pressure in the pipe.

Step 2:
Carefully withdraw the

flow meter stem from
the pipe until the sensor
head is fully inside the
hot tap adapter.

3.4 REMOVAL OF THE METER