ONICON D-1400 User Manual

Page 21

Advertising
background image

11451 Belcher Road South, Largo, FL 33773 • USA • Tel +1 (727) 447-6140 • Fax (727) 442-5699 • [email protected]
D-1400 Portable Metereing System Manual 05/14 - 0722-2

Page 21

SECTION 5.0: DIAGNOSTICS

5.1 DIAGNOSTICS

The ONICON D-100 display is programmable. Factory programmed settings provide rate and

total values in accordance with the customer’s application data. Refer to the D-100 calibration

sheet for a complete listing of factory settings. These settings may be reviewed and changed with

assistance from ONICON factory service personnel. The display is also equipped with diagnostic

indicator lights and self diagnostic test signals that confirm the operation of the microprocessor

and its input circuitry. Please contact the ONICON factory service personnel if any of the

diagnostic lights or test signals listed below indicates a potential problem with the operation of

the Btu meter.

5.1.1 Diagnostic Lights

Low Voltage Power Supply

Located on the power supply board inside the System-10 BTU Meter (refer to page A-3),

these two LEDs will be illuminated when the positive and negative five volt supplies

are present.

Liquid Flow

Located just above the power supply on the motherboard (refer to page A-1) is the flow

indicator LED. The LED will flash at a rate that is proportional to the liquid flow rate.

An unlit LED indicates no flow signal.

5.1.2 Flow Test Signals

Located on the right side of the motherboard (refer to A-1) immediately above the power

supply board is a three position slide switch used to test the flow input. When the switch

is in the top position, the input is connected to the flow meter. When the switch is in the

middle position, the input is connected to two test terminals used to apply a variable

frequency to simulate flow. When the switch is in the bottom position, the input is

connected to a 50/60 Hz signal that simulates a fixed flow rate. Refer to the calibration

certificate to determine the correct display reading when operating in the 50/60 Hz test

mode.

5.1.3 Ai3 & Ai4 Test Signals

As you are facing the component side of the processor board, the Ai3/Ai4 Test push button

is along the top edge of the board just to the left of the flat ribbon cable. It is labeled TEMP

TEST. When pressed, a fixed current will be applied to each sensor input. Both Ai3 and

Ai4 will produce an output equal to 37.5% of their programmed full scale output. This test

confirms the operation of the D-100 processor board input circuitry for Ai3 and Ai4.

IMPORTANT NOTE

The second line of the display will alternate between TEST MODE and the normal display of

engineering units and multipliers whenever the flow test switch is in the TEST or 60 Hz positions.

i

IMPORTANT NOTE

After operating for five minutes in either the TEST or 60 Hz mode, the displayed flow and energy

rate will be disabled, and the meter will report a zero flow rate and a zero energy rate to the

network. The meter will remain in this state until the switch is set to RUN.

i

Advertising