Figure 3, Figure 4 – Super Systems 9015 Series User Manual

Page 6

Advertising
background image

Super Systems Inc. Page 6 of 11

Series 9015 Operations Manual

Figure 3 - Editing Slave Instrument Settings

Figure 4 - Slave Instrument Setup Menu with Instrument 1 Configured

After you have completed the procedure above, “Instrument 1” under the “Slave

Instruments” menu will read “Instrument 1 [OK]” in the Parameter column (see

Figure

5

). It may take up to 10 seconds for the indication to change to “Instrument 1 [OK]”. (If

communications with the slave instrument are not working, “Instrument 1 [BAD]” will

be shown. If communications are being established, “Instrument 1 [N/A]??” will be

shown before the indication changes to “[OK]”.)

Advertising