7

Before starting the wiring of the instrument, check that the supply voltage available is the
same as marked on the label of the converter.

Terminal

Power supply AC

Power supply DC

11 Phase

DC

12 Neutral

DC

In case of DC power supply, the equipment is provided by a diode bridge that allows
connecting it regardless of the polarity .

3.2.2 Pulse

input

wiring

Terminal

1

Voltage output +5V

2

Contact

3

Ground

This input is compatible with all oscillating piston flowmeters series COVOL, as well as
with instruments that provide an output of the following types:

-

From 0 to VCC, where VCC = 5 … 24V

-

NPN open collector

-

Free potential contact

For those devices requiring be powered to 5VDC to generate output pulses, the MT03F
have this supply voltage in Terminal 1.

Free potential contact