Using Hydrosteel 6000

Before proceeding further, ensure that you are familiar with the user recommendations and the analyser
menu structure described below.

User recommendations
Hydrosteel 6000 is a robust unit with electrical and pneumatic design features to ensure that accidental
misuse does not represent a threat to the user or to the sensitive system components. However, we
recommend that you:


*

consult this manual or the service team at Ion Science when you are unsure of how to proceed.

*

avoid exposing any part of the system to corrosive media or liquids.

*

avoid exposing the system components to ambient temperatures that exceed 50

o

C for more than a

few hours.

*

when not in use, store all the system components and equipment in a cool dry place.


Analyser Operation Menu

Three front-panel keys
On the analyser front panel, the three keys

S (Spanner),
D (Data) and
O (On/Off)

have dual functions, according to
whether the analyser is in:
‘start-up mode’, for 90 s normalisation after the analyser is switched
on, or…
‘operating mode’, after normalisation.

Key

Start-up mode

Operating mode

S

adjust time, flux units, probe,

temperature scale

select zone, check calibration

D

handle data log

log data

O

turn off / zero

turn off / zero



The key options available are shown on the analyser display. After completing a function, the analyser
returns to start-up mode or operating mode, as appropriate. For example, in start-up mode, you can divert to
reset the clock because of a change to Summer Time and, on completion of this function, the analyser
automatically completes the normalising procedure.

For reference, a complete list of the key sequences, together with countdowns and available interrupt keys is
provided in Appendix B.







Ex

Ex

Ex98E2316

Eex ia IIC T4

-20 C< Ta< 60 C