Process input, D.ddd, dd.dd, ddd.d – Precision Digital PD603 User Manual

Page 7

Advertising
background image

Sabre P Model PD603 Process Meter

Instruction Manual

7

ENCLOSURE

1/8 DIN, high impact plastic, UL 94V-0, color: gray

MOUNTING

1/8 DIN panel cutout required. Two panel mounting brack-
et assemblies provided

TIGHTENING
TORQUE

Screw terminal connectors: 4.5 lb-in (0.5 Nm)

OVERALL
DIMENSIONS

2.45" x 4.68" x 4.19" (62 mm x 119 mm x 106 mm)
(H x W x D)

WEIGHT

8.5 oz (241 g)

WARRANTY

1 year parts & labor

EXTENDED
WARRANTY

1 or 2 years, refer to the Price List for details.

Process Input

INPUTS

Field selectable:
0-20, 4-20 mA, 0-5, 1-5, 0-10 V, or

10 V

ACCURACY

±0.05% FS ±1 count

FUNCTION

Linear

TEMPERATURE
DRIFT

50 PPM/

C from 0 to 65C ambient

DECIMAL POINT Up to three decimal places:

d.ddd, dd.dd, ddd.d,

or

dddd

CALIBRATION
RANGE

An Error message will appear if input 1 and input 2 signals
are too close together.

Input
Range

Minimum Span
Input 1 & Input 2

4-20 mA 0.40 mA
10 V

0.20 V

INPUT
IMPEDANCE

Voltage ranges: greater than 1 M

Current ranges: 50 - 100

(depending on resettable fuse impedance)

INPUT
OVERLOAD

Current input protected by resettable fuse.
Fuse resets automatically after fault is removed.

Advertising