Low flow operation verification procedure, Usa and canada only – Precision Medical HeliO2 Blenders User Manual

Page 17

Advertising
background image

Heliox / Oxygen Blender Service Manual

Page 15

LOW Flow Operation Verification Procedure

USA and CANADA ONLY

(50 psi / 3.45 bar MODELS)

S E Q #

DIAL

SET

O2%

OXYGEN PRESS

±1.0

HELIOX PRESS

±1.0

FLOWMETER

SET TO

lpm ±0.2

AUXILIARY

BLEED

FUNCTION

TARGET VALUE

ACTUAL

VALUE

psi

bar

psi

bar

1

ANY

50

3.45

50

3.45

closed

closed

leak

<2 psi / 2 MIN

*2

60

75

5.17

0

0

0

closed

back flow

<100 ml/min

*3

60

0

0

75

5.17

0

closed

back flow

<100 ml/min

4

21

50

3.45

50

3.45

3

open

end point

(±0.3) Source Value

5

40

50

3.45

50

3.45

3

open

set point

37.0%-43.0%

6

60

50

3.45

50

3.45

3

open

set point

57.0%-63.0%

7

80

50

3.45

50

3.45

3

open

set point

77.0%-83.0%

8

100

50

3.45

50

3.45

3

open

end point

(±0.3) Source Value

9

60

50

3.45

50

3.45

1

open

set point

57.0%-63.0%

10

60

60

4.14

67

4.62

1

open

set point

57.0%-63.0%

11

60

60

4.14

50

3.45

1

open

set point

57.0%-63.0%

12

60

50

3.45

Slowly

reduce to

30

Slowly

reduce to

2.07

3

closed

Alarm ON

30.0

± 2.0 psi

13

60

50

3.45

Slowly Increase until

alarm shuts off

3

closed

Alarm OFF

45.0 psi MAX

14

60

Slowly

Reduce to

30

Slowly

Reduce to

2.07

50

3.45

3

closed

Alarm ON

30.0

± 2.0 psi

15

60

Slowly Increase until

alarm shuts off

50

3.45

3

closed

Alarm OFF

45.0 psi MAX

16

60

50

3.45

50

3.45

MAX

closed

flow rate

30.0 lpm MIN

17

60

50

3.45

0

MAX

closed

flow rate

30.0 lpm MIN

18

60

0

50

3.45

MAX

closed

flow rate

30.0 lpm MIN

19

60

50

3.45

50

3.45

MAX

open

flow rate

30.0 lpm MIN

* Reference, Letter D. (Reverse Gas Flow Procedure)