Heliox / Oxygen Blender Service Manual

Page 16

II. HIGH FLOW (PM5500 Model)

Step 1: DISASSEMBLY

Tools Required

#2 Phillips Screwdriver

11/32 in. Nut Driver

#3 Phillips Screwdriver

5/32 in. Long Hex Key

1/2 in. Open End Wrench

Small Retaining Ring Pliers

Figure A

1. Rotate dial [1] to the 60 graduation.
2. Remove the two flat head screws [2] on each side of the

top cover [3].

3. Remove top cover by pulling upwards.

The cover will not come off unless the dial is at the 60
graduation







Figure B

4. Use a ½ in. open end wrench to unscrew and remove the

Heliox [4] and Oxygen [5] inlet assemblies and primary [9]
from bottom of the Blender.
Oxygen inlet has left hand threads.

5. If manifold outlet assembly is present, unscrew the stem

using a 5/32 hex key and holding the manifold outlet
assembly to the bottom of the Blender.

6. Remove the four flat head screws [2] from bottom cover

[6].

7. Remove bottom cover.