Precision Medical HeliO2 Blenders User Manual

Page 19

Advertising
background image

Heliox / Oxygen Blender Service Manual

Page 17

HIGH FLOW

Figure C

8. Remove dial [1] by pulling dial away from

manifold block [8].

9. Remove the primary [9] and auxiliary [10]

outlets (auxiliary contains Blue Muffler [38])
by using a ½ in. open end wrench to
unscrew.

10. Use Retaining Ring Pliers to unscrew and

remove audio alarm assembly [11] from the
bottom of the manifold block.

Figure D

11. Use small retaining ring pliers to remove retaining

rings [12] from each side of the alarm assembly
[13]. Push the alarm assembly through to remove
the assembly from manifold block.









Figure E

12. Using a 11/32 in. nut driver or socket, loosen

the nut [14] (ONLY two turns) holding the
knob guide [15] on the proportioning valve
assembly [16]. Slide knob guide assembly
from proportioning valve assembly shaft.
Slide resistance ring [17] from proportioning
valve assembly.

13. Remove phillips head screw [18] and washer

[19] from side of proportioning valve
assembly.

14. Replace knob guide assembly and tighten the

nut.

15. Pull knob guide assembly to remove

proportioning valve assembly from manifold
block. Again, loosen the nut (ONLY two turns) holding the knob guide to the proportioning valve
assembly, remove knob guide from proportioning valve assembly shaft.

Advertising