Cleaning / decontamination, Maintenance, Returns – Precision Medical PreSet Continuous Vacuum Regulator 0-200mmHg User Manual

Page 10

Advertising
background image

9

CONTINUOUS & CONTINUOUS INTERMITTENT

CLEANING / DECONTAMINATION

(As needed)

1. Attach a working Vacuum Regulator with a continuous regulated

mode to a minimum vacuum source of 15 inHg.

2. Mix cold disinfection/sterilization solution according to its

manufacturer’s directions.

3. Connect tubing as shown in Cleaning Illustration on previous page.

4. Turn the working Vacuum Regulator on to a continuous regulated mode.

5. Adjust the vacuum to a minimum of 120 mmHg.

6. Set the “Preset” Vacuum Regulator to be cleaned to the “REG.”

mode, and set at 100 mmHg.

7. Allow cold disinfection/sterilization solution to pass through Preset

Vacuum Regulator and collect in Suction Canister. Procedure should

continue for time recommended by the manufacturer of the cold

disinfection/sterilization solution for the desired level of disinfection or

sterilization.

8. Turn the Preset Vacuum Regulator to be cleaned to the “LINE” mode

(if applicable)

.

9. Allow remaining cold disinfection/sterilization solution to pass through

and collect in Suction Canister.

10. Set working Vacuum Regulator to its maximum vacuum setting.

11. Thoroughly dry the internal components by drawing maximum

vacuum through the cleaned Regulator for at least 30 seconds in

both “REG.” and “LINE” modes (if applicable).

NOTE: If it is not possible to pass cold disinfection/sterilization solution

through the Regulator, then the passageways are totally blocked

and DISASSEMBLY of the Regulator is required. Ensure to follow

the Biohazard protocol at your facility.

MAINTENANCE

Before each use, visually inspect Preset Vacuum Regulator for any sign

of damage. DO NOT USE if damaged.

RETURNS

Returned products require a Returned Goods Authorization (RGA) number,

contact Precision Medical, Inc. All returns must be packaged in sealed

containers to prevent damage. Precision Medical, Inc. will not be responsible

for goods damaged in transit. Refer to Precision Medical, Inc. Return Policy

available on the Internet, www.precisionmedical.com.
Note: Federal Law prohibits shipping of BioHazard materials Via U.S. mail.

Manuals available on our Website; www.precisionmedical.com