Caution, Preset vacuum regulator cleaning illustration – Precision Medical PreSet Continuous Vacuum Regulator 0-200mmHg User Manual

Page 9

Advertising
background image

8

CONTINUOUS & CONTINUOUS INTERMITTENT

Tell us how we are doing!

Visit us at www.precisionmedical.com

CAUTION

DO NOT autoclave or immerse in liquid. This will cause damage

to the Preset Vacuum Regulator and will void the warranty.

• If Preset Vacuum Regulator becomes internally contaminated,

warranty is voided, DO NOT send back to Precision Medical, Inc. for

repair. Follow the contaminated equipment protocol at your facility.

• Ensure all connections are tight and leak free.

• The Preset Vacuum Regulator contains magnetic, ferrous

material that may affect the results of an MRI.

PRESET VACUUM REGULATOR CLEANING ILLUSTRATION

(Cleaning/Decontamination Instructions on

page 9)

Preset Vacuum

Regulator

To Be Cleaned

Cold

Disinfection/

Sterilization

Solution

Suction

Tubing

Patient

Port

Back (Inlet) Connection

Suction

Collection

Canister

Vacuum

Port

Suction

Tubing

Working

Vacuum

Regulator