Specifications – Precision Medical PreSet Continuous Vacuum Regulator 0-200mmHg User Manual

Page 4

Advertising
background image

3

CONTINUOUS & CONTINUOUS INTERMITTENT

SPECIFICATIONS

CALIBRATED RANGE

P

S3000:

0 - 200 mmHg

PS3000HV: 0 - 300 mmHg

PS3100:

0 - 200 mmHg

PS3100HV: 0 - 300 mmHg

PS3300:

0 - 200 mmHg

PS3300HV: 0 - 300 mmHg

PS3400:

0 - 150 mmHg

PS3500:

0 - 150 mmHg

PS3600:

0 - 760 mmHg

GAUGE ACCURACY

Analog Gauge

±

5% of MAX calibrated range

Dual Analog/Digital Gauge

Digital Display

± 1% of Full Scale

Analog Gauge

± 5% of MAX calibrated range within ref. Indicator

INLET/OUTLET CONNECTIONS 1/8 NPT Female

MODES OF OPERATION

PS3000 Series

PS3500 Series

OFF

- No Vacuum

REG. - (Regulated) Provides an adjustable, continuous vacuum level

PS3100 Series

PS3600Series

OFF

- No Vacuum

REG. - (Regulated) Provides an adjustable, continuous vacuum level

LINE - Provides maximum, continuous vacuum from the vacuum source

PS3300 Series

PS3400 series

OFF - No Vacuum

REG. - (Regulated) Provides an adjustable, continuous vacuum level

INT. - Intermittent) Provides an adjustable vacuum level that cycles

between ON and OFF

FLOW RATES

PS3000, PS3100 & PS3300

REG MODE @ 200 mmHg

48 l/min

LINE MODE

55 l/min

INT. MODE (PS3300 Only)

10 l/min

PS3000HV, PS3100HV & PS3300HV

REG MODE @ 300 mmHg

50 l/min

LINE MODE

55 l/min

INT. MODE (PS3300HV Only)

10 l/min

PS3300 & PS3400

REG MODE @ 150 mmHg

50 l/min

INT. Mode (PS3300 only)

10 l/min

INT. MODE (PS3400 only)

3 l/min

PS3600

REG MODE @ 635 mmHg

71 l/min

LINE MODE

82 l/min

FLOW RATES ARE

OBTAINED WITH A VACUUM SOURCE OF 21” Hg

INTERMITTENT CYCLE TIME:

Factory set at sixteen (16) seconds ON

and eight (8) seconds OFF (Reference only)

Operating Environmental Limits:

0°F to 122°F (-18°C to 50°C)

Recommended Environmental Operating Limits: 55°F to 85°F (13°C to 29°C)
Storage Environmental Limits

Temperature Range:

-4°F to 140°F (-20°C to 60°C)

Humidity:

Max 95% Noncondensing

Battery:

3 Volt Lithium,

1

/

2

AA (Digital Vacuum Gauge Models ONLY)

Specifications are subject to change without prior notice.