7

CONTINUOUS & CONTINUOUS INTERMITTENT

ASSEMBLY INSTRUCTIONS

1. To assemble, perform the “DISASSEMBLY INSTRUCTIONS” in reverse order.

Note: • Ensure the tab on the Regulator Module is inserted into the internal

slot on the Selector Module at the 9 o’clock position.

• Ensure the tab on the Control Knob Assembly at the 12 o’clock

position is aligned with the outer circumference slot on the

Selector Module at the 12 o’clock position.

2. Lubricate all O-rings and cavities with Vacuum grease (part# 1775) supplied in

the Preset Vacuum Regulator Repair Kit.

3. Repeat steps 1 through 3 of “OPERATING INSTRUCTIONS”.

4. Prior to returning Preset Vacuum Regulator to service verify preset vacuum

level with gauge reading.

DISASSEMBLY INSTRUCTIONS

for PS3300 & PS3400 Series

(Reference “PARTS DESCRIPTION”)

1. Loosen the Set Screw (Item # 11) in Selector Knob.
2. Pull the Regulator Module/Control Knob Assembly (Item # 12) away from

case. (The Regulator Module (Item # 10) is threaded onto the Control Knob

Assembly. Note: These components must always be kept together as a

matched set.)

3. Remove the screws (Item # 2) from the back of the product.
4. Remove the Rear Case (Item # 14) by pulling away from product.
5. Remove screws (Item# 2) from the top of the Timing Module.
6. Remove the Timing Module (Item# 13) by pulling away from the Housing

Assembly (Item# 1).

7. Separate the Case Assembly (Item# 9) by pulling it away from the Housing

Assembly (Item# 1).

8. Remove the Selector Assembly (Item# 6) by pulling it away from the Housing

Assembly (Item# 1).

9. Remove Gauge by pulling vertically out of socket (Item# 3).

DISASSEMBLY INSTRUCTIONS

for PS3000, PS3100, PS3500

& PS3600 Series

(Reference “PARTS DESCRIPTION”)

1. Loosen the Set Screw (Item# 11) in Selector Knob.
2.

Pull the Regulator Module/Control Knob Assembly (Item # 12) away from case.

(The Regulator Module (Item # 10) is threaded onto the Control Knob Assembly.

Note

: These components must always be kept together as a matched set.)

3. Remove the screws (Item# 2) from the back of the Regulator.
4. Separate the Case Assembly (Item# 9) by pulling it away from the Housing

Assembly (Item# 1).

5. Remove the Selector Assembly (Item# 6) by pulling it away from the Housing

Assembly (Item# 1).

6. Remove Gauge by pulling vertically out of socket (Item# 3).

Advertising