RIGOL DS6000 Series User Manual

Page 32

Advertising
background image

RIGOL

2

RP1002C User’s Guide

Control/Indicator Description

Current flow symbol: The arrow shows
the positive direction of current flow
.

Zero adjustment: After connecting to
power and fixing the current probe to a
fixed position, rotate the knob to adjust
the probe output tothe proper DC offset. It
may also be used to offset a DC signal
component. Zeroing is not needed for AC
measurements unless your instrument
cannot isolate a DC component (if
present).

Range switch: Slide the switch from OFF
to either the 50 mV/A or 500mV/A range.
When 50 mV/A or 500mV/A is selected,
the probe is turned on, and the green
battery indicator (ON) lights.

Power Supply indicator: The green
power supply indicator lights when the
probe is turned on. For more information,
please refer to To Install the Battery.

Overload indicator: The red overload
indicator lights and flicks continuously if
the measured signal is greater than the
maximum range of the selected range
capacity.



Advertising
This manual is related to the following products: