RIGOL

RP1002C User’s Guide

1

Basic Operation


Before using the current probe, the battery or specified AC power
adapter must be installed. For the battery installation instructions,
refer to To Install the Battery.

CAUTION

Do not clamp the probe onto circuits with voltages
greater than 600 VAC. Otherwise, personal injury or
damage to the probe may result.


CAUTION

Always connect the RP1002C current probe output to the
instrument before clamping onto the circuit under test.

Connection between the Current Probe and
Oscilloscope


1 Set the coupling mode of the oscilloscope to DC and connect

the BNC interface of the RP1002C current probe to the
oscilloscope input via dual-BNC coaxial cable.

2 Move the Range switch to the 50 mV/A or 500 mV/A position

to turn on the RP1002C current probe. At this point, the green
battery indicator lights.

3 Use the zero adjustment knob to adjust the probe to the

proper DC offset.