Chapter 8 Counter

RIGOL

DG4000 Series User’s Guide

8-3

To Set the Counter


You need to configure appropriate parameters for the counter.

1. TrigSens

Set the trigger sensitivity of the measurement system. The default is 50% and
the range available is from 0% to 100%.

Press TrigSens. Use the numeric keyboard to input the desired value and
select the unit “%” from the pop-up menu.

2. TrigLevel

Set the trigger level of the measurement system. The system triggers and
gets the measurement readings when the input signal reaches the specified
trigger level. The default is 0V and the range available is from -2.5V to 2.5V.

Press TrigLevel. Use the numeric keyboard to input the desired value and
select the desired unit (V or mV) from the pop-up menu.

3. Impedance

Set the input impedance to “50Ω” or “HighZ” and the default is “HighZ”.

4. Coupling

Set the coupling mode of the input signal to “AC” or “DC” and the default is
“AC”.

5. Atten

Set the attenuation coefficient of the input signal to “X1” or “X10”.

6. HFReject

High-frequency reject can be used to filter out the high-frequency
components and improve the measurement accuracy in low-frequency signal
measurement.

Press HFReject to enable or disable the high-frequency reject function.