RIGOL DM3068 User Manual

Page 39

Advertising
background image

RIGOL

DM3068 Quick Guide

V

Input Terminal Protection Limit

The protection limit applies to input terminals:

1 Main input (HI and LO) terminals

HI and LO terminals are used for Voltage, Resistance, Capacitance, Continuity,
Frequency and Diodes measurements and should be used under the following
two conditions:
1) HI-LO protection limit: at most 1000 VDC or 750 VAC, this is also the

maximum measurable voltage. The limit can be expressed as 1000 Vpk.

2) LO-ground protection limit: at most 500 Vpk (relative) is allowed to float at

LO terminal with safety.

Since the HI terminal holds a maximum protection of 1000 Vpk relative to the
ground, the sum of the “float” and measured voltages cannot exceed 1000 Vpk.

2 Sampling (HI Sense and LO Sense/200 mA) terminals

HI Sense and LO Sense/200 mA terminals are used for 4-Wire Resistance
measurement and should be used under the following two conditions:
1) HI Sense-LO Sense/200 mA protection limit: 200 Vpk.
2) LO Sense/200 mA-LO protection limits: 0.5 Vpk. The current input fuse

on the rear panel provides the current passing through LO Sense/200
mA
up to 500 mA protection.

3 Current input (10 A and Sense/200 mA) terminals

10 A and LO terminals are used for current measurements of 2 A and 10 A. The
maximum current which goes through the 10 A terminal is limited to 10 A by
the internal fuse. LO Sense/200 mA and LO terminals are used for current
measurements ranging from 200 µA to 200 mA. The maximum current which go
through the LO Sense/200 mA terminal is limited to 500 mA by the internal
fuse.

NOTE:
In order to prevent the fuse from blowing out and protect the
multimeter, please use the current input terminals according to the
following requirements:

1) Do not connect the 10 A and LO Sense/200 mA input terminals into the

current measuring circuit at the same time.

2) Only use 10 A and LO terminals for measurements when the measured

Advertising