Quick Guide for DSA1000 Series

37

Troubleshooting

To help you solve commonly encountered problems we have listed some
typical issues with their respective solutions. If the problems persist contact
RIGOL and prepare your device information (SystemInformation
System Info).

1. The screen still dark (no display) after power on:

(1) Check if the fan is running:

If running, the connection of the internal screen cable may be

loose.

If not, the instrument has failed to power up, please refer to step (2)

or (3).

(2) If the battery (China only) is used for power supply:

Check whether the battery is in good condition and whether it is

correctly installed.

Check whether the battery status indicator

at the front panel

goes on; if not, the battery contains no electricity and please
charge the battery.

(3) If AC power is used for power supply:

Check if the power supply has been connected correctly and power

switch has been turned on.

Check if the power fuse is burned. If a new fuse needs to be

installed, please select one with these specifications: 250V AC, T2A;
5 mm×20 mm.

2. Unresponsive keys or wrong response

(1) Press all the keys on the front panel to make sure that each of them is

normal after power on.

(2) Press SystemSelf-TestKey Test to see if all keys are working

properly.

(3) If a key is not working, do not disassemble the instrument by yourself