A.9 environmental limits, A.10 environmental effects, A.11 hazardous area classifications – Emerson 2400S User Manual

Page 37: Environmental limits, A.10, Environmental effects, A.11, Hazardous area classifications

Advertising
background image

Installation Manual

33

Dimensions and Specifications

Re
tu
rn
P

o

li

cy

In

de
x

Spec

ifications

A.9

Environmental limits

A.10

Environmental effects

A.11

Hazardous area classifications

Ambient temperature limits

Operating and storage: –40 to +140 °F (–40 to +60 °C)

Below –4 °F (–20 °C), LCD responsiveness decreases and LCD may become
difficult to read. Above 131 °F (55 °C), some darkening of the LCD panel might
occur.

ATEX requires limiting ambient temperature to below 131 °F (55 °C).

Humidity limits

5 to 95% relative humidity, non-condensing at 140 °F (60 °C)

Vibration limits

Meets IEC68.2.6, endurance sweep, 5 to 2000 Hz, 50 sweep cycles at 1.0 g

All models

EMI effects

Complies with electromagnetic compatibility directive 2004/108/EC per
EN 61326 industrial

Conforms to NAMUR NE21 Version: 10.02.2004

Model 2400S Analog only

Ambient temperature effect

On mA output: ±0.005% of span per °C

All models

CSA

(1)

C-US

(1) CSA is a Canadian approvals agency that provides approvals accepted both in the U.S.A. and in Canada.

Class I Division 2 Groups A, B, C, D

Class II Division 2 Groups F and G

ATEX

(2)

(2) ATEX is a European directive.

Analog or PROFIBUS-DP

II 3G EEx n A C II T5

II 3D IP66/IP67 T70°C

DeviceNet

II 3G Ex nA II T5

II 3D IP66/IP67 T70°C