Specifications, Body mawp, Str14g 500 psig (35 bar) – Honeywell STR12D User Manual

Page 5: Str17g 3000psig (207 bar), Str14a 500 psia (35 bar), Operating conditions – all models

Advertising
background image

34-ST-03-64

Page

5

Specifications

Operating Conditions – All Models

Parameter

Reference Condition

Rated Condition

Operative Limits

Transportation and

Storage

°C °F °C °F °C °F °C °F

Ambient Temperature *

25 ±1

77 ±2

–55 to 90 –67 to 194

Humidity %

RH

10 to 55

0 to 100

0 to 100

0 to 100

Maximum Allowable

Working Pressure (MAWP)

MAWP is minimum of Body Rating or Seal Rating (See Model Selection Guide for Seal

MAWP)

Body

MAWP

STR12D 2500 psig (172 bar) Bolted Process Heads Table I _ _ A

STR13D 2500 psig (172 bar) Bolted Process Heads Table I _ _ A
STR12D

1450 psig (100 bar) All Welded Process Heads Table I _ _ C

STR13D 1450 psig (100 bar) All Welded Process Heads Table I _ _ C

STR14G 500 psig (35 bar)
STR17G 3000psig (207 bar)

STR14A 500 psia (35 bar).

Vacuum Region - Minimum

Pressure

mmHg

absolute

Atmospheric (See Figure 15 for vacuum limitations.)

Supply Voltage, Current,

and Load Resistance

Voltage Range:

10.8 to 42.4 Vdc at terminals

Current Range:

3.0 to 21.8 mA

Load Resistance:

0 to 1440 ohms (as shown in Figure 16)

* Ambient Temperature Limit is a function of Process Interface Temperature. (See Figure 13.)