MICROLAB

®

Carousel User Guide (June 2003)

16

For the 8-position platter, align the #7 position line with the center line of
the index pointer.

For the 10-position platter, align the #9 position line with the center line
of the index pointer. Then type 18 into the keyboard resolution field, and
hit the right-arrow key once.

6. Click OK. A window appears asking you to “Reset instrument

now”. Unplug the carousel, wait 5 seconds, and plug the carousel
back in. Wait another 5 seconds, and click OK.

Figure 15:

Calibration alignment