Troubleshooting – Hamilton Electronics MICROLAB 8534-01 User Manual

Page 27

Advertising
background image

MICROLAB

®

Carousel User Guide (June 2003)

25

Troubleshooting

In the event of an error during run time, click on the Help button (if
available). The message should provide some clues to the cause of the error
and a possible remedy. If not, try to reproduce the error, record the error code
and message, and notify Hamilton Company Technical Service. See
“Contacting Hamilton Company” section on page 29. In addition, table 3
lists the most common error conditions for the carousel and the associated
corrective actions.

Table 3.

Common error conditions

Error code

Possible cause

Corrective action

Ox5E 8001

Unable to initialize

Contact Hamilton Service

Ox5E 8002

Unable to initialize

Contact Hamilton Service

Ox5E 8003

Motion not detected

Bad sensor; contact
Hamilton Service

Ox5E 8004

Motor stalled

Remove obstacles. Avoid
moving platter by hand
during motion.

Advertising