Table 4 wiring the sensor at terminal block j5, Figure 8 hard-wiring the digital gateway, Table 4 – Hach 6120118 User Manual

Page 15: Figure 8, Installation, Disconnect power from probe

Advertising
background image

15

Installation

Figure 8

Hard-wiring the Digital Gateway

3.2.2 Connecting the Digital Gateway to a sc100 Controller in a Hazardous Location

DANGER
The sc100 and certain versions of the sensor are suitable for use in Class 1,
Division 2, Groups A, B, C, D Hazardous Locations . See Control Drawing 58600-78
in the sc100 Controller Manual, Cat. No. 58600-18 for acceptable sensor versions
and installation requirements.

DANGER
Le sc100 et certaines versions du capteur peuvent être utilisés dans des endroits
dangereux de la Classe 1, Division 2, Groupes A, B, C, D. Reportez-vous au schéma
de contrôle 58600-78 du Manuel du contrôleur sc100, Réf. 58600-18 pour connaître
les versions des capteurs admises et les conditions d'installation.

DANGER
Explosion hazard. Do not connect or disconnect equipment unless power has been
switched off or the area is known to be non-hazardous.

DANGER
Risque d’explosion. Couper le courant ou s’assurer que l’emplacement est designe
non dangereux avant de replacer le aucon composant.

Table 4 Wiring the Sensor at Terminal Block J5

Terminal Number

Terminal Designation

Wire Color

1

Data (+)

Blue

2

Data (–)

White

3

Service Request

No Connection

4

+12 V dc

Brown

5

Circuit Common

Black

6

Shield

Shield (grey wire in existing quick disconnect fitting)

1

1

+ DATA

+ DATA

+ OUT 2

+ OUT 2

– DATA

– OUT 2

SERVICE REQUEST

SERVICE REQUEST

SHIELD/CHASSIS GND

+ V

+ V

+ OUT 1

+ OUT 1

GND

– OUT 1

2

2

3

3

4

4

5

5

6

PROBES

ANAL

OG OUTP

UTS

ANAL

OG OUTP

UTS

PCB

CONNECTOR

PCB

CONNECTOR

FIELD WIRING

INSULATION MUST

BE RATED TO

80° C MINIMUM

FIELD WIRING

INSULATION MUST

BE RATED TO

80° C MINIMUM

DANGER - EXPLOSION HAZARD

DANGER - RISQUE D'EXPLOSION

DO NOT DISCONNECT WHILE CIRCUIT IS LIVE

UNLESS AREA IS KNOWN TO BE NON-HAZARDOUS.

NE PAS DEBRANCHER TANT QUE LE EST SOUS

TENSION, A MONIS QU'IL NE S'AGISSE D'UN

EMPLACEMENT NON-DANGEROUX

NC

NC

NC

COM

COM

COM

NO

F1

F2

NO

NO

RELAY 3

RELAY 2

RELAY 1

1

1

+ DATA

+ DATA

+ OUT 2

+ OUT 2

– DATA

– OUT 2

SERVICE REQUEST

SERVICE REQUEST

SHIELD/CHASSIS GND

+ V

+ V

+ OUT 1

+ OUT 1

GND

– OUT 1

2

2

3

3

4

4

5

5

6

PROBES

ANAL

OG OUTP

UTS

ANAL

OG OUTP

UTS

PCB

CONNECTOR

PCB

CONNECTOR

FIELD WIRING

INSULATION MUST

BE RATED TO

80° C MINIMUM

FIELD WIRING

INSULATION MUST

BE RATED TO

80° C MINIMUM

J1

J2

J4

NETWORK

INTERFACE

CARD

J3

U5

U9

S1

DANGER - EXPLOSION HAZARD

DANGER - RISQUE D'EXPLOSION

DO NOT DISCONNECT WHILE CIRCUIT IS LIVE

UNLESS AREA IS KNOWN TO BE NON-HAZARDOUS.

NE PAS DEBRANCHER TANT QUE LE EST SOUS

TENSION, A MONIS QU'IL NE S'AGISSE D'UN

EMPLACEMENT NON-DANGEROUX

1

1

+ DATA

+ DATA

+ OUT 2

+ OUT 2

– DATA

– OUT 2

SERVICE REQUEST

SERVICE REQUEST

SHIELD/CHASSIS GND

+ V

+ V

+ OUT 1

+ OUT 1

GND

– OUT 1

2

2

3

3

4

4

5

5

6

PROBES

ANAL

OG OUTP

UTS

ANAL

OG OUTP

UTS

PCB

CONNECTOR

PCB

CONNECTOR

FIELD WIRING

INSULATION MUST

BE RATED TO

80° C MINIMUM

FIELD WIRING

INSULATION MUST

BE RATED TO

80° C MINIMUM

DANGER - EXPLOSION HAZARD

DANGER - RISQUE D'EXPLOSION

DO NOT DISCONNECT WHILE CIRCUIT IS LIVE

UNLESS AREA IS KNOWN TO BE NON-HAZARDOUS.

NE PAS DEBRANCHER TANT QUE LE EST SOUS

TENSION, A MONIS QU'IL NE S'AGISSE D'UN

EMPLACEMENT NON-DANGEROUX

NC

NC

NC

COM

COM

COM

NO

F1

F2

NO

NO

RELAY 3

RELAY 2

RELAY 1

1

1

+ DATA

+ DATA

+ OUT 2

+ OUT 2

– DATA

– OUT 2

SERVICE REQUEST

SERVICE REQUEST

SHIELD/CHASSIS GND

+ V

+ V

+ OUT 1

+ OUT 1

GND

– OUT 1

2

2

3

3

4

4

5

5

6

PROBES

ANAL

OG OUTP

UTS

ANAL

OG OUTP

UTS

PCB

CONNECTOR

PCB

CONNECTOR

FIELD WIRING

INSULATION MUST

BE RATED TO

80° C MINIMUM

FIELD WIRING

INSULATION MUST

BE RATED TO

80° C MINIMUM

J1

J2

J4

NETWORK

INTERFACE

CARD

J3

U5

U9

S1

DANGER - EXPLOSION HAZARD

DANGER - RISQUE D'EXPLOSION

DO NOT DISCONNECT WHILE CIRCUIT IS LIVE

UNLESS AREA IS KNOWN TO BE NON-HAZARDOUS.

NE PAS DEBRANCHER TANT QUE LE EST SOUS

TENSION, A MONIS QU'IL NE S'AGISSE D'UN

EMPLACEMENT NON-DANGEROUX

J6

J5

J5

J5

J5

Disconnect
Power

From Probe