Standard operations – Hach DR 5000 User Manual

Page 43

Advertising
background image

43

Standard Operations

Store Off/On

With the Store On setting, all measurement data are stored automatically. With the Store Off
setting, no measurement data are stored.

% Trans/Conc/Abs

To switch to % transmittance, concentration or absorbance readings

Send Data icon /
Send Data

To send Data to a printer, computer or USB memory stick (Type A)

Timer icon

This functions as a stopwatch. It helps to ensure that the steps of an analysis are correctly timed
(e.g. reaction times, wait times, etc., can be exactly specified). When the specified time has
elapsed, an acoustic signal is emitted. The use of the timer has no influence on the measurement
program.

Reading Mode

Single Reading Mode: A reading is only displayed after a measurement has been carried out
(press Read; standard setting) (see

section 5.4.4.1 on page 45

).

Continuous Reading Mode: After the zero measurement, all readings are displayed
automatically and continuously (see

section 5.4.4.2 on page 45

).

Dilution Factor Off/On

A corrective dilution factor can be entered in order to take account of certain properties.
The number entered at the dilution factor prompt will be multiplied by the result to compensate for
the adjustment. For example, if the sample has been diluted by a factor of 2, enter 2. The default
setting of the dilution factor is turned off.
Note: When a dilution is in effect, the dilution icon will appear on the display.

Standard Addition

This enables the accuracy of the measurements to be checked. The (working) procedure for a
test parameter contains a detailed explanation of how to use this function.

Standard Adjust

The (working) procedure for a test parameter indicates whether a standard adjustment is
necessary and, if so, how to proceed.

Chemical Form

Some of the stored tests/methods allow to select the chemical form and the measuring range.

Reagent Blank

Some of the stored tests/methods include the "Reagent Blank" function. This enables the reagent
blank value to be added to or subtracted from, the subsequent readings. The reagent blank value
shifts the calibration curve along the y-axis, without changing the shape or gradient of the curve.
The effect corresponds to a y-axis intercept of the calibration straight line. This is made clear by
the following equation: Concentration = [(Conc. factor) * Abs] – (reagent blank value).

Save as User Program

To store the selected parameters as a User Program, see

section 6.1 on page 61

.

Recall Data

Call up saved measurement data, wavelength scans or time courses, see

section 5.3 on

page 35

.

Instrument Setup

Basic data of the instrument, see

section 5.2 on page 23

.

Table 4 Stored programs options (continued)

Options

Description

Advertising