1 complete a barcode 13 mm test/vial, 2 select the measuring range, 3 select the chemical evaluation form – Hach DR 5000 User Manual

Page 51

Advertising
background image

51

Standard Operations

5.5.1 Complete a barcode 13 mm test/vial

Note: The cell compartment must be closed before a measurement is
made!

1. The instrument needs to be at the Main Menu.

2. Prepare the barcode 13 mm cuvette/vial in accordance with the

working procedure and insert the cuvette/vial in cell
compartment #1 and close the cell compartment.

When a coded cuvette/vial is placed in cell compartment #1
(

Figure 7

), the corresponding measurement program is

automatically activated in the Main Menu.

The measurement is started automatically and the results are
displayed.

To evaluate other cuvette/vial tests and other parameters, insert the
prepared cuvette/vial into the cell compartment and read the result.

5.5.2 Select the measuring range

Some tests can be used for different measuring ranges. After the
sample cuvette/vial has been inserted, a list of the different
measuring ranges is displayed.

Select the required measuring range by highlighting the appropriate
line.

Press Start Permanent if this measuring range is to apply to all
subsequent measurements.

Changing the standard setting
In the result display, press Options>More>Permanent: On. The
key will change to Permanent: Off.

5.5.3 Select the chemical evaluation form

The chemical form of the test result of a number of parameters can
be selected individually.

In the result display, press the unit (e.g. mg/L) or the chemical
representation of the evaluation form (e.g. PO

4

3–

–P). A list of

possible evaluation forms is displayed, from which the required
form can be selected. Press OK to confirm.

Another way of changing the standard setting is: