Advanced operations – Hach DR 5000 User Manual

Page 89

Advertising
background image

89

Advanced Operations

Reading mode:

1. To highlight the required mode, start by pressing Reading

mode.

2. Select the required mode, then press OK, then Return to return

to the result display.

6.4.2 Take single wavelength measurements (single reading)

1. Insert the blank cuvette/cell into the cuvette/sample cell holder.

Press Zero.

Note: The Read key is only active after the zero measurement has
been completed.

2. Insert the sample cuvette/cell into the cuvette/sample cell

holder. Press Read.

3. For data storage, see

section 5.3.2 on page 38

.

6.4.3 Take single wavelength measurements (continuous readings)

1. Insert the blank cuvette/cell into the cuvette/sample cell holder.

Press Zero.

Note: In the reading mode ”Continuous“ only the Zero key is shown to
start the reading. The reading sequence is started automatically.

2. Insert the sample cuvette/cell into the cuvette/sample cell

holder.

3. Press Options and then the Store icon to store the displayed

data in the Data Log.

Note: For data storage, see

section 5.3.2 on page 38

.