Advanced Operations

Program Number:
Specific test number, with which the program can subsequently be
called up from the selection list in the User Programs menu or the
Favorites menu.

2. Use the alphanumeric keypad to enter a program number

between 950 and 999. The lowest available number appears
automatically.

3. Press OK.

Note: If the program number is already assigned to another user
program, a message appears, asking whether the existing program
should be replaced. Press OK to overwrite the existing program.

Program Name:

4. Use the alphanumeric keypad to enter a program name. The

name can be a maximum of 28 characters long.

5. Press Back to go back to the previous program point or press

Next to continue with the input of the program data.

Program Type:

6. Select the required option (

Table 8

) and press Next.

7. If the Single Wavelength (

section 6.1.1.1 on page 63

) or Multi

Wavelength (

section 6.1.1.2 on page 64

) is selected, define the

unit, wavelength, absorbance formula, wavelength

λx,

concentration factor Kx, concentration resolution, chemical
form and calibration equation.
More information on Free Programming parameters see

section 6.1.2 on page 72

.

Table 8 Program descriptions

Program Type

Description

Single Wavelength

Measurements at a defined wavelength

Multi Wavelength

In the Multi Wavelength mode, absorbance values can be measured at up to four wavelengths
and the results can be mathematically processed to obtain sums, differences and relationships.

Free Programming

This is an advanced form of programming for original user-developed methods. In Free
Programming, the user defines the measurement process, variables and calculations involved in
obtaining a reading.