Local Operation

Menu

4

4-9

Menu | Setup — Limits

The Menu | Setup | Limits menu is used to setup all of the limits in the system. The

pressure limits can be used to protect the device-under-test (DUT) from overpressure.

gld12.bmp

Figure 4-10. Menu | Setup | Limits - Menu

High Limit

High limit is the user-defined Maximum pressure limit. This is often set just over the

full-scale pressure of the device under test (DUT) in order to protect the DUT. The

Calibrator will then prevent the pressure from exceeding the high limit. If the user enters

a value greater than the user defined limit, the unit will not accept the value and it will

generate an error code informing the user that they have entered a set-point that is greater

than the high pressure limit. In control mode, if for any reason the pressure exceeds the

user defined limit, it will change into the measure mode and again display an error

message to the operator informing them of the error that occurred.
This value can be edited using the rotary knob to highlight the High Limit. Then enter the

new value using the keypad. This will show up in the scratch pad on the lower portion of

the screen. Press the Enter key and the value for the high limit will be updated to the

value that was typed into the scratchpad.

Low Limit

User defined Minimum pressure limit. This is the same as the high limit except it is to

protect the DUT from low pressure limits.

Auto Vent

User defined maximum pressure that the Calibrator can reach prior to venting the test

port to atmosphere.