Fluke RUSKA 7350 User Manual

Page 36

Advertising
background image

RUSKA 7350

Users Manual

4-12

Step Size

In addition to entering the pressure set-point by the key-pad, the pressure set-point can be

changed using the step function. The step function is used primarily when the operator is

taking pressure steps in equal pressure increments. The size of the pressure step is user

defined. From the Main Menu, press Menu [F6], Setup [F2], and then User [F2]. Using

the rotary knob, move the cursor to highlight Step Size. Enter the desired step size in the

current pressure units and press ENTER.

Bar Graph Maximum

The bar graph on the Main Menu screen can be scaled to match the device under test by

setting the full scale value of the bar graph.

Ready Tolerance

A Ready indication is generated when in control mode and the measured pressure is

reading within this ready tolerance value. When running an internal program, the ready

indicator is shown when the measured pressure is within the tolerance stored in the

program.

Gas Head Pressure Correction

The term “head height” refers to the vertical distance between the sensing element in the

device under test and the Calibrator’s pressure reference plane. Once the user inputs the

head height and selects air or nitrogen, the Calibrator automatically corrects for head

pressure.
1. The Pressure Reference Line on the Calibrator is defined as the bottom of the color

display where the display and the front panel join. This provides the reference plane

against which the device under test (DUT) pressure is measured.

2. Determine the vertical distance between the Calibrator Pressure Reference Line and

the reference plane of the device under test.

3. Press the Unit key found in the keypad and press in/mm [F1] to select either inches

(in) or millimeters (mm) for the head height entry.

4. The head height is set from the Menu | Setup | User. From the Main Menu (press

PREVIOUS until the Main Menu appears), press Menu [F6

⎢Setup [F2] ⎢User [F2].

5. Press Medium [F3] to select either “Air” or “Nitrogen”. The selected units will be

highlighted.

6. Press Position [F4] to select if the position of the Device Under Test is either

“Above” or “Below” the 7350. The selected units will be highlighted and appear in

the Gas Head description.

7. Use the rotary knob to highlight the label “Gas Head”.
8. Use the numeric keypad to enter the height in the selected unit.
9. Press ENTER to accept the entry.

Atmosphere

The nominal barometric pressure value for the location that the Calibrator is utilized

should be entered for atmosphere. On tare mode instruments, this number is used to

calculate the density of the test port pressure medium and to correct for head pressure

variances between the reference port of the Calibrator and the DUT.