Fluke RUSKA 7350 User Manual

Page 57

Advertising
background image

Remote Operation

Device Messages

5

5-5

:MODE?

calibration

edit

enabled?

:MODE ON¦OFF¦1¦0

enable calibration edit

(Cal.

button

required)

DISP
:ENABle ON¦OFF¦1¦0

turns front panel display on/off

:TEXT <string>

displays message on front panel

:BGRaph <number>

sets bar graph maximum

OUTPut
:STATe ON¦OFF¦1¦0

off=MEASure, on=CONTrol

:STATe?

returns 0=Measure or 1=Control

:MODE

MEASure¦CONTrol¦VENT

sets

mode

:MODE?

returns

mode

string

PROGram
:CATalog?

returns list of defined programs

[SELected]
:DEFine <program block>

Define program

press1,

toler1,

dwell1,

max1,

press2,

toler2,

..

:DEFine?

read

program

definition

:DELete

[:SELected]

deletes current program

:ALL

deletes

all

programs

:NAME <program name>

select current program

:STATe RUN|PAUSe|STOP|CONTinue

set program state

:STATe?

`

read

program

state

:CONFigure

:RECall

restore

saved

configuration

:SAVE

save

current

configuration

SENSE
[:PRESSure]
[:RESolution] <number>

set pressure display resolution

:AUTO <boolean>|ONCE

return to default resolution

:MODE?

returns ABSOLUTE or TARE

:RANGE

[:UPPer]

returns RUSKA 7350 full scale value in

units

:LOWer

returns

0

:REFerence

[:HEIGht] <number>

set head height

:SGRavity <numbers>

set specific gravity

[SOURCE]
[:PRESsure]
[:LEVel]
[:IMMediate]

[:AMPLitude]

<number>

sets

pressure

setpoint

[:AMPLitude]?

read

pressure

setpoint

:MODE FIXed|LIST

set source parameter set

:TOLerance <number>

specifies output tolerance

:SLEW <number>

set slew rate

:CONTrol <number>

set control band

Advertising