B-1

Appendix B

Summary of Error Messages

Summary of Error Messages

Negative error numbers are from the Standard Commands for Programmable Instruments

(Version 1991.0).

Table B-1. Summary of Error Messages

Value

Description and Corrective Action

0 No

Error.

-103

Invalid Separator. Check punctuation in the SCPI command.

-104

Data Type. The type of parameter data is incorrect. Check for numeric versus string
data.

-109

Missing Parameter. No valid parameter was found for the SCPI command.

-110

Command Header. The command name is not valid.

-113

Command Unknown. The command specified does not exist.

-114

Header Suffix. The numeric suffix for the command name is out of range.

-201

Remote Required. Remote command conflicted with local interface. Return local
interface to main Menu.

-221

Settings Conflict. The command could not be executed due to the current state of
the Calibrator. Some commands cannot be executed while a program, self-test, or
calibration is in progress.

-222

Out of Range. The value is not within the valid range. For pressures, check your
high and low limits.

-281

Cannot create program. Program memory is full.

-282

Illegal Program Name. The name specified is not valid or does not exist.

-284

Program Currently Running. The command cannot be executed while a program is
running.

-285

Program Syntax Error. The syntax of the program definition is not correct.