Local Operation

Menu

4

4-25

gld26.bmp

Figure 4-24. Menu | Test | Sweep - Menu

4. Press Run [F2] to initiate the test.
5. To pause the program, press Pause [F3]. Pause will now be highlighted and the

Calibrator will continue controlling to the current set-point. The Calibrator will

maintain control of the current set-point until further instructions from the user. Press

Continue [F4] to resume the program.

6. To stop the program, press Stop [F5]. The program will stop running but the

Calibrator will continue controlling to the current set-point.

Menu | Test — Self Test

The Calibrator can perform electrical and pneumatic self test to assist in trouble shooting

potential problems. Please refer to Chapter 6, Maintenance, for more details.

Note

In order to perform a pneumatic self test, the supply port must be connected
to a properly set supply pressure and the test port not open to atmosphere.
For absolute units, a vacuum supply must also be connected to the
exhaust port.

Menu | Test — Remote Test

The Menu | Test | Remote section of the system is used to perform a number of the

diagnostic functions on the remote interface. It can be used to display the transmitted and

received messages that are sent across the interface. This can be a powerful tool to assist

in identifying the source of communication problems.
The Menu | Test | Remote menu displays information from the IEEE-488 interface.

Pressing the Serial 1 [F2] or Serial 2 [F3] keys will display information from the

serial ports.