Real-Ear Measurements

97

5.6.1.1 To perform the coupler measurement

The coupler measurement part of the RECD is saved into the analyzer’s permanent memory until
another measurement is stored on top of it. This saves a lot of time because it means you only have
to perform the real-ear part of the RECD for each client instead of performing both the coupler and
the real-ear measurements.

Note: Steps 1-7 are optional but good to do in order to get the most accurate RECD.

1. Press F1 from the Main Coupler screen (not the Target 2-cc screen) to enter the setup menu.

2. Press F4 — CAL MIC(S).

3. Attach the calibration clip to the larger coupler microphone so that the metal tube hangs off

the very tip of the microphone. See Figure 5.6.1.1A.

Coupler

mic

HA-2

Coupler

To 3A

insert

earphone

25 mm

(I.D. 1.93 mm)

Figure 5.6.1.1A—Attaching the two microphones Figure 5.6.1.1B—Coupling the coupler microphone

together

to the insert earphone.

4. Thread probe tube connected to the probe microphone through tube so that the probe tube

sticks out over the grill of the coupler microphone.

5. Put assembly in sound chamber and close chamber door.

6. Press F6 – COMPEN PROBE. This will measure and store the differences between the two

microphones.

7. Press F1 – EXIT.