Chapter 7: Telecoil Testing

Telecoil testing is available on serial number 940000 and above, manufactured in September, 1994
and later. Testing with the Telewand is available with software version 3.60 and above.

7.1 Setup with the Telecoil Board

1. Set up the hearing aid in the usual way, by connecting it to the correct coupler and insert the

test microphone. You may use either a normal hearing aid battery or a battery substitution
pill to power the aid.

2. Place the telecoil board close to the FP40 as shown in Figure 7.1A.

CAUTION

ELECTRICAL SHOCK HAZARD.

DO NOT REMOVE INSTRUMENT COVER.

REFER SERVICE TO QUALIFIED PERSONNEL.

EXTERNAL

POWER

ON

OFF

HEADPHONES

EXTERNAL

SPEAKER

VIDEO MONITOR

SERIAL

INTERFACE

FRYE

ELECTRONICS, INC.

®

EXTERNAL

SPEAKER

TELECOIL BOARD

Figure 7.1A Telecoil setup

3. Connect the telecoil to the back of the FP40 at the jack marked external speaker.

4. Press F1 – MENU from the Main Coupler screen in order to enter the Setup Menu.

5. Use the AMPLITUDE and FREQUENCY knobs to select MAIN F2 under FUNCTION KEY

DEFIN.

6. Press START/STOP repeatedly to select TELECOIL.