Real-Ear Measurements

77

Top View

Speaker

(45 degrees)

12"

Reference Mic

Probe Mic

12"

Reference Mic

Probe Mic

Front View

Speaker

(45 degrees)

Figure 5.1.2.1—Positioning of the sound field speaker for real-ear measurements

5.1.2.2 To place the earhook and reference microphone

1. Place the wedge-style earhook on the client’s ear. Alternately, place an earhook (without a

wedge) on the client’s ear, and place the Velcro headband around the client’s head.

2. Attach the reference microphone, facing for-

ward, on the wedge of the earhook, directly
above the ear to be tested. The reference micro-
phone should be as close to the head as pos-
sible. See Figure 5.1.2.2.

3. Adjust the round Velcro pad on the wedge style

earhook by turning it, so that the inside pad fits
firmly against the client’s neck.

Figure 5.1.2.2—Placing the reference

microphone