2640A/2645A NetDAQ

Users Manual

6-22

4.

Verify Accuracy Configure the Decade Resistance Source for the output
values below and verify the Spy window measurement is between the
minimum and maximum values. Change the channel 1 range as required (see
Step 2).

Resistance Range

Decade Resistor

Minimum Reading

Maximum Reading

300

Short Circuit (Zero)

0

0.05

300

290

289.861

290.139

3 k

Short Circuit (Zero)

0

0.5

3 k

2.9 k

2.89849 k

2.90137 k

30 k

29 k

28.9834 k

29.0166 k

300 k

290 k

289.621 k

290.379 k

3 M

2.9 M

2.89146 M

2.90854 M

The resistance accuracy in this table makes allowance for up to 0.01% decade resistance
tolerance.

Resistance Range

5700A

Minimum Reading

Maximum Reading

300

Short Circuit (Zero)

0

0.030

300

190

189.912

190.088

3 k

Short Circuit (Zero)

0

0.30

3 k

1.9 k

1.89912 k

1.90088 k

30 k

19 k

18.9893 k

19.0107 k

300 k

190 k

189.750 k

190.250 k

3 M

1.9 M

1.89425 M

1.90575 M

5.

Close Spy Window.