Displaying the Totalizer Status

2-29.

Perform the procedure in Figure 2-16 to display the instrument Totalizer status.
The instrument updates the Totalizer display once per second. (See Figure 2-15
for examples.) To clear the Totalizer count, cycle the instrument power. You can
also configure NetDAQ Logger to clear the Totalizer count when it starts logging.

•

Maximum Totalizer Count The Totalizer counts switch closures or voltage
transitions with a maximum count of 4,294,967,295.

•

Totalizer Status Display The instrument displays the Totalizer status in two
five-digit segments; high (HI) and low (LO). Using the maximum count of
4,294,967,295 as an example, the HI five-digit segment would display 42949
and the LO five-digit segment 67295.

DIO

Press the DIO (Digital I/O) key to view the Totalizer status. You can view the
Totalizer status whether the instrument is scanning or not scanning.

Press an up/down arrow key to advance to the Totalizer display. The 10

digits display in a five-digit tot:HI count and five-digit tot:LO count.

DIO

Press the DIO key again to exit. NetDAQ Logger also provides a display of the
Totalizer status.

Figure 2-16. Displaying the Totalizer Status