A-10 – Fluke NetDAQ 2645A User Manual

Page 220

Advertising
background image

2640A/2645A NetDAQ

Users Manual

A-10

Table A-12. 2640A AC Voltage General Specifications (cont)

Specification

Characteristic

Maximum Input Voltage

The lesser voltage of:
300V ac rms from any terminal on channels 1 and
11 to earth.

150V ac rms from any terminal on channels 2
through 10, and 12 through 20 to earth.

300V ac rms from any terminal on channels 1 and
11 to any other terminal.

150V ac rms from any terminal on channels 2
through 10 and 12 through 20 to any other input
terminal.

Maximum Volt-Hertz Product

2x10

6

Volt-Hertz product on any range, normal

mode input.

1x10

6

Volt-Hertz product on any range, common

mode input.

Temperature Coefficient

Linear interpolation between 2 applicable points for
temperatures between 28

º

C and 60

º

C, or -10

º

C

and 18

º

C, e.g., if the applicable specification at

28

º

C is 2% and the specification at 60

º

C is 3%,

then the specification at 40

º

C is

(3%-2%)x(40-28)/(60-28)+2%=2.375%.

Accuracy at -20

°

C

Multiply the -10

°

C to + 60

°

C accuracy specification

by 2. After 1 hour warm-up. For accuracy between
-10

°

C and -20

°

C, interpolate linearly.

DC Component Error

The presence of a dc voltage will cause an
indeterminate error in the reading of the ac voltage
on the input.

Table A-13. 2640A AC Voltage Range and Resolution Specifications

Range

Resolution

Minimum Input for

Slow

Fast

Rate Accuracy

Full Scale

±

30,000

±

3,000

300 mV

10 µV

100 µV

20 mV

3V

100 µV

1 mV

200 mV

30V

1 mV

10 mV

2V

150/300V

10 mV

100 mV

20V

Note 300V range applies to channels 1 and 11 only.