Maintenance

Introduction

6

6-3

Introduction

6-1.

Maintenance for the 2640A/2645A instruments is limited to self-test error code
explanations, cleaning, fuse replacement, performance test, calibration, and a
listing of replacement parts. A Service Manual (PN 942615) is available for
purchase.

Self-Test Diagnostics and Error Codes

6-2.

Self-test diagnostics are performed each time the instrument is powered up. Any
errors encountered during this initial 5-second period are reported on the front
panel, as shown below.

If you encounter an error code, refer to Table 6-1 for a brief description of the
error. For all errors, try cycling the instrument power. If the error persists and you
intend to repair the instrument yourself, refer to the Service Manual, PN 942615.
Otherwise, package the instrument securely (using the original container, if
available), and mail it to the nearest Fluke Service Center. Include a description of
the problem. Fluke assumes no responsibility for damage in transit.

Table 6-1. Self-Test Error Codes

Error Code

Error Description

1

Boot ROM Checksum Error

2

Instrument ROM Checksum Error

3

Internal RAM Test Failed

4

Display Power-Up Test Failed

5

Display Not Responding

6

Calibration Constants Corrupted

7

A-D Not Responding

8

A-D Self-Test Failed

9

A-D Zero Offset Test Failed

10

A-D Reference Balance Test Failed

11

A-D Overload Detection Test Failed

12

A-D Open Thermocouple Detection Test Failed

13

Communication Parameters (e.g., baud, IP address) Corrupted

14

Ethernet Address Corrupted

15

Internal RAM Constants Corrupted

16

Ethernet Chip Test Failed

17

Uncalibrated or Calibration Incomplete