Volts ac accuracy test, Volts ac accuracy test 6-11 – Fluke NetDAQ 2645A User Manual

Page 185

Advertising
background image

Maintenance

Performance Test

6

6-13

3.

Verify Accuracy Configure the 5700A for the output values below and
verify the Spy window measurement is between the minimum and maximum
values. Change the channel 1 range as required (see Step 1).

Volts DC Range

5700A Output

Minimum Reading

Maximum Reading

90 mV

Short Circuit (Zero)

-0.000023V

+0.000023V

90 mV

+90 mV

+0.089965V

+0.090035V

90 mV

-90 mV

-0.090035V

-0.089965V

300 mV

Short Circuit (Zero)

-0.00005V

+0.00005V

300 mV

+300 mV

+0.29991V

+0.30009V

300 mV

-300 mV

-0.30009V

-0.29991V

3V

Short Circuit (Zero)

-0.0004V

+0.0004V

3V

+3V

+2.9992V

+3.0008V

3V

-3V

-3.0008V

-2.9992V

30V

Short Circuit (Zero)

-0.005V

+0.005V

30V

+30V

+29.991V

+30.009V

30V

-30V

-30.009V

-29.991V

50V

Short Circuit (Zero)

-0.04V

+0.04V

50V

+50V

+49.95V

+50.05V

50V

-50V

-50.05V

-49.95V

4.

Close Spy Window.

Volts AC Accuracy Test

6-11.

Complete the following procedure to test the accuracy of the volts ac function for
both the 2640A and 2645A. Measurement accuracy applies to all channels, not
just the channel used for the test.

1.

Configure Channel 1 for Volts AC In NetDAQ Logger, configure channel 1
for volts ac, 300 mV range. (See “Configuring Analog Channel Functions" in
Chapter 3 of this manual.)

2.

Open Spy Window Select the Spy command from the Utilities menu. Select
channel 0101 (instrument 01, channel 01) from the Channel list. Click OK to
open the Spy window.